Description
Use of the app
The Scan Controller TCP Settings app is provided to set network settings for REVOLON and DISS6 scan controllers. It is needed for the initial hardware installation on the microscope or lab network, and for subsequent changes to the network.
Please make sure any TCP change is compatible with the existing microscope or lab network!
For connection to the scan controller, client apps will also require these TCP settings. If a client app fails to connect to the device, please double check app setting with the TCP settings on the display panel on the front of the hardware scan controller.

Start
Components At start,of the appUI
will
Status
Connection status will be given in the text line under the window title bar. This may bebe:
- [...] connected over USB
- [...] connected over TCP
- No USB Device found
where [...] may be 'REVOLON' or 'DISS6', depending on what hardware is found.
For further information see the Connection page in this manual.
HelpPlease click
Settings
These
Address
The TCP/IP address of the device.
This field is disabled if DHCP is enabled.
For IP address set in the device, please always read the TCP/IP address from the touch display of the hardware scan controller.
MaskThe subnet mask of the device. Default setting is 255.255.255.0.
This field is disabled if no connection is made to the device.
Protocol
A number of portsPorts used by the hardware scan controller to communicate on the network. Default settings are usually sufficient, unless these ports are already reserved on the local network for other communication.
| Port | Description |
| SUP | Port for SEM Upgrade Protocol |
| DSCP | Port for DISS Scan Comissioning Protocol |
| DIIP | Port for DISS Information Instruction Protocol |
| DIDP | Port for DISS Image Data Protocol |
| DVDP | Port for DISS Video Data Protocol |
Turns the TCP/IPTCP communication on or off.
Default settingsetting:
6 = checkbox deactivated
DHCP EnabledTurns the DHCP client on or off. Default setting is off.
Connect TCPIf a USB connection is made at app start, then the 'Connect TCP' button is disabled. If a USB connection is not made at app start, then the 'Connect TCP' button is available to connect to the hardware scan controller using a TCP/IP address. For further information, see the ConnectionsConnection page in this manual.
To apply settings, please click on 'Apply & Close', which will prompt to switch the device off and on again, to apply the settings.
Exit app
Use the close button of the app main window to exit the app without applying new settings - this will keep existing device settings.